Acceptance tests at radiographic and fluoroscopic X-ray units have to be carried out according to IEC 61223-3-1.
The constancy test for general direct radiography has to be done according to IEC 61223-2-11 (DIN 6868-3).
The test parameters are
- Dose measurements:
- Radiation output from the X-ray source assembly
- Radiation input to the image reception area (alternatively: measurement of the optical density at specific points of a radiogram)
- Geometric characteristics
- Resolution of high-contrast detail
- Variation in optical density throughout a radiogram.
To carry out the test the following equipment is necessary:
- Dosimeter, e.g. DOSIMAX plus (the use of a double dosemeter – DOSIMAX plus duo or PID duo with two detectors - makes it possible to measure both dose values simultaneously, see Double dosemeter philosophy, page 6-8)
- Test device ETR1 with centering tube
- Densitometer
Additionally the film screen contact should be tested whenever a problem is suspected or at least once a year (ISO 4090) using the Test Device FFA 4090 R.
In the case of Fluoroscopic Installations using X-ray image intensifiers and during the recording of the exit screen of the X-ray image intensifier, the constancy test, in accordance with German Standard 6868-4, the following parameters has to be tested by using ETR1 and a dosemeter
- Optical density in RBV recordings
- Spatial resolution
- Contrast
- Size of the useful beam field
- Dose rate or dose
Digital Subtraction Angiographic (DSA) X-ray installations must be tested for specific functions in accordance with the standard IEC 61223-3-3 resp. DIN 6868-8, therefore a special Test Tool is required for this, the RÖVI-8. This test device fulfils the special requirements of this standards and comprises the following components:
- an acrylic attenuator (150 x 150 x 57 mm)
- a dynamic copper step wedge ranging from 0.2 to 1.4 mm, ascending with
- an additional contrast step from 1.4 mm Cu to 0.2 mm
- a moveable insert with Al stripes for vascular simulation
Tests must be carried out for the following:
- Dose per image
- Dynamic range and contrast sensitivity
- Spatial resolution
- Artifacts
